Laser-induced fluorescence measurement of plasma ion temperatures: Corrections for power saturation
Source abstract
Abstract
When using high-power pulsed tunable dye lasers to measure plasma ion temperatures, it is important to attenuate the laser intensity. The temperature is found from the Doppler broadening of a spectral line. This may be obscured by saturation broadening, an instrumental effect encountered when too much laser intensity is used. Three useful experimental methods for determining the optimum pulsed laser intensity are found from a semiclassical atomic physics model. As an example, an Ar II transition pumped by a 1-GHz bandwidth laser is examined. The fluorescence linewidth of room-temperature ions broadens from 1.50 to 2.87 GHz when the homogeneous laser intensity is increased from 50 kW/m2 to 5 MW/m2.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Laser-induced spectroscopy and plasma · Spectroscopy and Laser Applications · Laser Design and Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Physics and Astronomy, The University of Iowa, Iowa City, Iowa 52242-1479