The molecular drag gauge as a calibration standard
Source abstract
Abstract
The National Institute of Standards and Technology has used molecular drag gauges for six years as transfer standards in the high-vacuum range, 1×10−4 to 1×10−1 Pa. We report on the experience gained with these gauges and, in particular, on their long- and short-term calibration stability, on factors affecting accuracy, on the predictability of the effective accommodation coefficient, and on factors affecting the stability of the offset correction.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Atomic and Subatomic Physics Research · Radioactive Decay and Measurement Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Center for Basic Standards, National Institute of Standards and Technology, Gaithersburg, Maryland 20899