The National Bureau of Standards primary high-vacuum standard
Source abstract
Abstract
The theory, design, operating procedures, and estimated errors are discussed for an orifice-flow-type pressure standard currently in use at the National Bureau of Standards. This standard is used to define pressures between 10−7 and 10−1 Pa. Including the uncertainty of the flowmeter, the estimated total uncertainty varies from 2.6% at the highest pressures, to 1.4% at midrange, and 8.2% at the lowest pressures. Representative calibration results are presented for four different types of hot-cathode ionization gauges.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Advanced Sensor Technologies Research · Flow Measurement and Analysis
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
National Bureau of Standards, Center for Basic Standards, Gaithersburg, Maryland 20899
References