Vacuum-calibration apparatus with pressure down to 10−10 Pa
Source abstract
Abstract
The vacuum-calibration apparatus is mainly composed of the extreme high-vacuum (XHV) system, ultrahigh vacuum system, and separated-flow system. The ultimate pressure of 7.9×10−10 Pa was obtained in the XHV calibration chamber by combining the magnetically levitated turbomolecular pump and nonevaporable getter pump (NEGP). The separated-flow method was used to extend the lower limit of vacuum-gauge calibration to 10−10 Pa. The uncertainty at lower limit was reduced when taking inert gases as test gases because NEGP has no pumping speed for inert gases. For this apparatus, the combined standard uncertainties were estimated to range from 1.5% at 10−4 Pa to 3.5% at 10−10 Pa.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Advanced Sensor Technologies Research · Radioactive Decay and Measurement Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Lanzhou Institute of Physics National Key Laboratory of Vacuum and Cryogenics Technology and Physics, , Lanzhou 730000, China
References (9)
Show source-provided references
1. J. Vac. Sci. Technol. A · 1997 · 15 · 239510.1116/1.580754 ↗
2. J. Vac. Sci. Technol. A · 1997 · 15 · 75310.1116/1.580815 ↗
3. 1980 · 255
4. Vacuum · 1999 · 53 · 29110.1016/S0042-207X(98)00373-X ↗
5. Metrologia · 1999 · 36 · 49310.1088/0026-1394/36/6/2 ↗
6. MAPAN-J. Metrology Soc. India · 2007 · 22 · 39
7. J. Vac. Sci. Technol. · 2007 · 26 · 92
8. Vacuum · 2007 · 44 · 1