journal article·1995-05-01

Precision shell characterization using radial averaging of x-ray images

Richard B. Stephens

Publisher page

Abstract

Radially averaging the image of a circular object makes full use of structural information contained in the photograph and optimizes interface detection and location. Measurement of shell dimensions from x-radiograph plates are limited by film graininess and camera resolution. These noise sources make edge determination using single line-outs very uncertain. The radial average process, by coherently adding radial line-outs from the entire circumference, greatly reduces that noise. Our system can detect interfaces between plastic layers with only slight differences in x-ray opacity and locate surface radii with an uncertainty of ≊0.2 μm. This capability has been created with a desktop computer and slight modification to a freeware image processing program, and careful system design.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Optical measurement and interference techniques · Medical Imaging Techniques and Applications · Welding Techniques and Residual Stresses

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

General Atomics, P.O. Box 85608, San Diego, California 92186-9784

References

Reference metadata was not supplied for this record.