Precision shell characterization using radial averaging of x-ray images
Source abstract
Abstract
Radially averaging the image of a circular object makes full use of structural information contained in the photograph and optimizes interface detection and location. Measurement of shell dimensions from x-radiograph plates are limited by film graininess and camera resolution. These noise sources make edge determination using single line-outs very uncertain. The radial average process, by coherently adding radial line-outs from the entire circumference, greatly reduces that noise. Our system can detect interfaces between plastic layers with only slight differences in x-ray opacity and locate surface radii with an uncertainty of ≊0.2 μm. This capability has been created with a desktop computer and slight modification to a freeware image processing program, and careful system design.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Optical measurement and interference techniques · Medical Imaging Techniques and Applications · Welding Techniques and Residual Stresses
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
General Atomics, P.O. Box 85608, San Diego, California 92186-9784
References