journal article·1991-09-01

New x-ray limit measurements for extractor gauges

Fumio Watanabe

Publisher page

Abstract

A new x-ray limit measurement method for extractor gauges has been developed using modulation of the hemispherical ion reflector. The measurement is possible from pressures as high as the 10−8 mbar range. X-ray limits for commercial extractor gauges have been examined for various emission currents and different gauge port geometries. The scatter in the x-ray limits measured is about 30% and the effect of gauge port geometry is more than 70%.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · X-ray Spectroscopy and Fluorescence Analysis · Nuclear Physics and Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Tsukuba Laboratory, Sukegawa Electric Co., Ltd., Toyoshima Building, 2157-1 Kamiyokoba, Tsukuba, Ibaraki 305, Japan

References

Reference metadata was not supplied for this record.