journal article·1987-07-01

Mass-dependent effects of channel electron multipliers in residual gas analyzers

Noreen R. Reagan, Louis C. Frees, John W. Gray

Publisher page

Abstract

Residual gas analyzers (RGA’s) are increasingly being used to obtain quantitative information about vacuum processes. The gain for an ion [electron multiplier (EM) output current divided by the Faraday cup (FC) output current] determined by experiment for a combination FC/EM sensor was not constant for increasing mass. The gain for ions of different masses was also found to be dependent on the chemical nature of the ion, reflecting the conversion efficiency of incident ions to secondary electrons [U. Fehn, Int. J. Mass Spectrom. Ion Phys. 15, 391 (1974); 21, 1 (1976)]. Allowance for these effects makes quantitative RGA measurements in the EM mode of the sensor more reliable. Methods for determining a correction factor are discussed.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Analytical Chemistry and Sensors · Scientific Measurement and Uncertainty Evaluation · Electrochemical Analysis and Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Inficon Leybold–Heraeus, Inc., East Syracuse, New York 13057

Galileo Electro-Optics Corp., Galileo Park, Sturbridge, Massachusetts 01518