Elimination of serious artifacts in temperature programmed desorption spectroscopy
Source abstract
Abstract
Regurgitation and wall displacement processes are often encountered in temperature programmed desorption measurements. Such effects can lead to misleading interpretations of the surface reaction and desorption products. A simple and useful pumping scheme is described which overcomes this problem by using a cryogenically cooled titanium sublimation pump (TSP) for differential pumping of the mass spectrometer. It is demonstrated that minor wall displacement effects (mainly CO, H2O, and H2) are present in temperature programmed desorption spectroscopy and are insensitive to the mode of differential pumping. In comparison to ion pumping, however, serious methane (CH4) regurgitation and displacement effects can be completely eliminated when the TSP is employed.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Advanced Chemical Physics Studies · Mass Spectrometry Techniques and Applications · Spectroscopy and Laser Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Chemistry, Surface Science Center, University of Pittsburgh, Pittsburgh, Pennsylvania 15260