journal article·1994-09-01

Increased precision in strain measurement of diamond by microRaman spectroscopy

W. Brock Alexander, Paul H. Holloway, Joseph Simmons, Romulo Ochoa

Publisher page

Abstract

MicroRaman spectroscopy has been used to measure stress in diamond with improved precision. Stress was induced by rf sputter deposited tungsten films of 1- and 9-μm thickness. Preliminary data showed a systematic error in the diamond Raman peak position of about 0.05 cm−1/scan, which led to inaccurate and imprecise stress measurement. A new technique, using a reference 5520-Å krypton line only 10.5 cm−1 from the Raman diamond line, results in an order of magnitude increase in measurement precision. Using this calibration, diamond peak position could be measured with a precision of ±0.03 cm−1 which corresponds to ±11 MPa of stress. This is an easy and inexpensive means to improve stress measurements in diamond by microRaman spectroscopy.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Diamond and Carbon-based Materials Research · Advanced Surface Polishing Techniques · Metal and Thin Film Mechanics

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Materials Science and Engineering, University of Florida, Gainesville, Florida 32611

Trenton State College, Trenton, New Jersey 08625

References

Reference metadata was not supplied for this record.