Zero stability of spinning rotor vacuum gages
Source abstract
Abstract
Spinning rotor vacuum gages measure pressure by determining the rate of slowing of a magnetically suspended spinning ball over and above the slowing caused by a pressure independent residual drag. For accurate measurement in the high vacuum range, this residual drag must be determined and subtracted as an offset correction. The stability of this residual drag, temperature induced changes of the ball’s moment of inertia, vibration, and random measurement noise will determine the limits of stability and hence, the lowest usable pressure of the gage. Selected balls in a quiet, stable environment have demonstrated instabilities as low as ±10−6 Pa (10−8 Torr) equivalent nitrogen pressure. However, instabilities as large as two orders of magnitude greater can occur. Examples are given of different types of instabilities and guidelines are presented for minimizing many of the sources of instability.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Fluid dynamics and aerodynamics studies · Tribology and Lubrication Engineering · Geomagnetism and Paleomagnetism Studies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Center for Basic Standards, National Bureau of Standards, Gaithersburg, Maryland 20899
References