Method for correcting peak overlaps in quantitative Auger electron spectroscopy of Cr-containing oxides
Source abstract
Abstract
This paper provides a new method for improving the quantitative accuracy of Auger electron spectroscopy (AES) when measuring the surface composition of mixed metal corrosion oxides that contain chromium (Cr). Without this correction, the Cr Auger peak intensity (in the conventional first derivative display mode) is attenuated by its overlap with the oxygen (O) Auger line, leading to significant relative errors (∼30%–70% too low) in the Cr content. A peak fitting method was used to derive basis spectra in the direct mode display (before differentiation) and simulate a wide range of O to Cr ratios. A mathematical function was derived to define a Cr intensity correction factor, as a function of the O/Cr ratio. This method can be conveniently applied to data sets containing many spectra, such as depth profiles. Once corrected, the composition depth profiles obtained by AES are consistent with x-ray photoelectron spectroscopy profiles obtained from similar areas. The approach presented can also be used to generate correction factors for overlapping peaks in the Auger spectra of other material systems.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Electron and X-Ray Spectroscopy Techniques · X-ray Spectroscopy and Fluorescence Analysis · Corrosion Behavior and Inhibition
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Naval Nuclear Laboratory , P.O. Box 1072, Schenectady, New York 12301-1072
References (7)
Show source-provided references
1. 1995
2. Seah · 1983 · 189
3. J. Vac. Sci. Technol. A · 1990 · 8 · 73510.1116/1.576956 ↗
4. Corros. Sci. · 1984 · 24 · 13710.1016/0010-938X(84)90042-8 ↗
5. J. Vac. Sci. Technol. A · 2008 · 26 · 65010.1116/1.2908433 ↗
6. Vacuum · 1989 · 39 · 43910.1016/0042-207X(89)90266-2 ↗
7. Anal. Chem. · 1982 · 54 · 251110.1021/ac00251a024 ↗