Absolute calibration of small gas amounts
Source abstract
Abstract
A relatively simple but accurate calibration method for small gas amounts in the UHV region using ‘‘synthetic flash desorption spectra’’ is described. Well defined gas amounts are introduced into the vacuum chamber from a special glass vessel. The particle number in the glass vessel is obtained using a spinning rotor gas friction gauge for absolute pressure measurements. Furthermore, secondary standards can be produced and calibrated which allow a quick and easy absolute determination of the gas influx into the vacuum system. Desorption spectra obtained from calibrated tungsten filaments are used as secondary standards. As an example for the application of the methods described the saturation coverage and the initial sticking coefficient for hydrogen on various single-crystal surfaces have been determined. Further applications are the determination of the effective pumping speed in the vacuum chamber and the determination of the conductance of small tubes in the molecular flow regime. Additionally, the calibration of ionization gauges with spinning rotor gauges is discussed with respect to gas mixtures.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Phase Equilibria and Thermodynamics · Spectroscopy and Laser Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Institut für Festkörperphysik, Technische Universität Graz, Petersgasse 16, 8010 Graz, Austria
References