Thermal noise in vacuum scanning tunneling microscopy at zero bias voltage
Source abstract
Abstract
At typical operating conditions of scanning tunneling microscopy (STM) the current noise is dominated by 1/f noise the nature of which is still unclear. However at zero bias and zero average current thermal noise corresponding to the equivalent tunneling conductivity σ is found. Within the experimental limits it agrees well with the Nyquist formula d/dω〈I2 〉=4kTσ, where I is the tunneling current, ω is the angular frequency, k is Boltzmann’s constant and T is the absolute temperature. Since the thermal current noise scales linearly with σ it has the same exponential distance dependence as the tunneling current and can be used to adjust the distance between tip and sample. The performance of such a ‘‘scanning noise microscope’’ based on that principle is compared to the conventional STM technique for the example of a polycrystalline silver surface in UHV. It is shown that this method can also be applied at ambient pressure. This novel technique might be advantageous whenever the effect of the tunneling current on the sample should be avoided, i.e., very delicate samples, electrolytic environment, etc.