Calibration of molecular drag vacuum gages
Source abstract
Abstract
In spinning-ball molecular drag gages, the logarithm of the rotational frequency varies linearly with the time integral of the applied pressure under conditions of free molecular motion. An orifice flow standard for calibrating these devices has been developed that passes a measured quantity of gas through the calibration chamber in typically 30 min. The time integral of the calibration pressure can be calculated with an estimated uncertainty less than 1%. A number of gage balls operated in conjunction with three controllers have been calibrated for N2, Ar, and He using this system. Conditions and procedures affecting the performance characteristics of these gages have been investigated, including magnetization of the ball, baking and other conditioning of its surface, and degradation of gas purity by surface contaminants in an unbaked flow system.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Radioactive Decay and Measurement Techniques · Gas Dynamics and Kinetic Theory
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Catholic University of America, Washington, D. C. 20064 Center for Absolute Physical Quantities, National Bureau of Standards, Washington, D. C. 20234
References