A system for vacuum gauge calibration using the comparison technique
Source abstract
Abstract
A system for calibrating vacuum gauges in the range 10−4 to 102 Pa with an estimated uncertainty of ±8% is described. The factors affecting the design of the calibration vessel are discussed and the performance of the system enumerated. The reference gauges used are triode ionization gauges and a capacitance manometer. The reproducibilities of the reference gauges are estimated together with the other factors which contribute to the total uncertainty attached to a calibration performed on the system.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Radioactive Decay and Measurement Techniques · Advanced Frequency and Time Standards
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
National Physical Laboratory, Teddington, Middlesex TW11 OLW, United Kingdom