journal article·1996-05-01

Enhanced ignition of cold cathode gauges through the use of radioactive isotopes

K. M. Welch, L. A. Smart, R. J. Todd

Publisher page

Abstract

The Relativistic Heavy Ion Collider (RHIC) at Brookhaven National Laboratory comprises vacuum systems that range in pressure from 10−4 to 10−11 Torr. For cost savings associated with commonality, it is desirable to use one ultrahigh vacuum (UHV)-compatible high vacuum gauge throughout the accelerator. Work was done with radioactive isotopes to enhance ignition of inverted magnetron gauges at very low pressures (i.e., <10−10 Torr). Tests were initially conducted using cobalt-coated copper washers with a nickel flash to prevent oxidation. Solid nickel washers were also activated and tested. Problems with these sources will be discussed. Finally, an americium source was tested. The UHV-compatible americium sources yielded average ignition times of about 10 min at 4×10−11 Torr. The addition of the americium source produced a gauge suitable for operation at UHV pressures, and as a consequence for all of the RHIC.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Magnetic confinement fusion research · Radioactive Decay and Measurement Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Brookhaven National Laboratory, Upton, New York 11973-5000

References

Reference metadata was not supplied for this record.