Cryopump vacuum recovery after pumping Ar and H2
Source abstract
Abstract
For many vacuum processes the rate of recovery to a specific base pressure after gas injection is used as a criterion for determining the cleanliness of the vacuum system. When a cryopump is used, a long recovery time does not necessarily mean that undesirable gas species are present, because the recovery time of the cryopump itself increases as it becomes loaded with gases. This paper reports on experimental measurements of recovery time for APD-8 Cryopumps after flowing Ar and H2 at different rates for extended periods of time. Tests were run to measure the effect of using a new supercarbon versus standard charcoal and to see the effect of the cryotrapping of H2 by Ar without the use of charcoal. The recovery time after injecting Ar correlates with the time for the surface temperature of the solid Ar to return to base temperature. The recovery time for H2 is a function of the amount of H2 that has been pumped, cryopanel temperature, type of charcoal, gas flow rate, H2 concentration, and sequence of gas injection.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Superconducting Materials and Applications · Spacecraft and Cryogenic Technologies · Heat Transfer and Boiling Studies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
APD Cryogenics Inc, Science Center, Allentown, Pennsylvania 18103
References