journal article·1996-09-01

Influence of envelope geometry on the sensitivity of ‘‘nude’’ ionization gauges

Albert R. Filippelli

Publisher page

Abstract

This article presents the results of some measurements of the influence of envelope size and shape on the N2 sensitivity of two nominally identical ‘‘nude’’ extractor ionization gauges, and three common ‘‘nude’’ versions of the Bayard–Alpert ionization gauge. Measurements were made over the pressure range 10−7–10−1 Pa, using typical gauge operating parameters and with vacuum chamber and gauge envelope at ground potential. Sensitivity values corresponding to the three different envelopes used in this work differed by as much as a factor of 2 (comparing maximum value to minimum value for each gauge) for the Bayard–Alpert gauges. For the extractor gauges, the differences were as large as 7%.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Mass Spectrometry Techniques and Applications · Radioactive Decay and Measurement Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Thermophysics Division, Chemical Science and Technology Laboratory, National Institute of Standards and Technology, Gaithersburg, Maryland 20899

References

Reference metadata was not supplied for this record.