journal article·1985-11-01

Ion current modulation of a residual gas analyzer

F. Watanabe, H. Ishimaru

Publisher page

Abstract

A residual gas analyzer (RGA) is described with a modified quadrupole mass filter. This change permits a separate analysis of ions created in the gas from those created by electron impact in the RGA. Electrically, the electron energies are modulated and the ion current is measured in a phase-lock loop. Experimental results on H+ and O+ indicate that the spurious electron-stimulated-ion-desorption peaks are eliminated.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Mass Spectrometry Techniques and Applications · Electrostatic Discharge in Electronics · Microfluidic and Capillary Electrophoresis Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Sakura-no-seibo Gakuin High School, Ohgida, Noda-machi, Fukushima-shi, 960, Japan

National Laboratory for High Energy Physics, Oho-machi, Tsukuba-gun, Ibaraki-ken, 350, Japan

References

Reference metadata was not supplied for this record.