Ion current modulation of a residual gas analyzer
Source abstract
Abstract
A residual gas analyzer (RGA) is described with a modified quadrupole mass filter. This change permits a separate analysis of ions created in the gas from those created by electron impact in the RGA. Electrically, the electron energies are modulated and the ion current is measured in a phase-lock loop. Experimental results on H+ and O+ indicate that the spurious electron-stimulated-ion-desorption peaks are eliminated.