journal article·1991-05-01Open access

The measurement of cathode evaporation rate by quartz crystal microbalance

Zhang-qi Cheng, Chong-rou Wu, Ming-sheng Ni

Abstract

The quartz crystal microbalance (QCM) is used to measure the thickness of deposited materials in vacuum. This paper presents a method for measuring cathode evaporation rate by the QCM. The cathode and quartz crystals are mounted in a vacuum tube which is evacuated to 1×10−5 Pa. To overcome thermal frequency shifts, a thermistor is used to control the heater current to maintain a constant temperature. The frequency stability of the QCM is better than 0.3 Hz, so that the detectable mass of the evaporated material is 3.5×10−9 g/cm2, which corresponds to 3×10−2 monolayer of barium. The evaporation rates for (Ba, Sr, Ca) O cathodes and scandate cathodes were measured by this QCM. The measured rates were about 2×10−7 g/cm2 min (at 1210 K) and 1.5×10−8 g/cm2 min (at 1273 K) respectively, which are in agreement with the data by other authors. Calculation of the evaporation rate for small flat cathodes and cylindical cathodes are also given.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Acoustic Wave Resonator Technologies · Gas Sensing Nanomaterials and Sensors · Advanced Sensor Technologies Research

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Electronic Engineering, Southeast University, Nanjing, Peoples Republic of China

References

Reference metadata was not supplied for this record.