Sputter ion pumps are nowadays available for operating at pressures up to 10−3 Torr. Problems arise when measuring pumping speed in the high pressure range (namely, 10−5–10−3 Torr) where the usual Fisher–Mommsen test dome is inapplicable. As a matter of fact assumptions on which it is based are not verified at such high pressures. In the present work a method is described for measuring pumping speeds in the high pressure range (10−6–10−3 Torr). Examples of application of this method to different pumps are given. Pumping speed plots are compared with the ones obtained with the Fisher–Mommsen test dome. In the 10−6–10−4 Torr range, where these two methods overlap, pumping speeds obtained by both methods are in fairly good agreement. By combining the two methods it is possible to characterize pumping speeds in the whole operating range.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Analytical Chemistry and Sensors · Mass Spectrometry Techniques and Applications · Acoustic Wave Resonator Technologies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Varian S.p.A., via F. lli Varian, 54-10040 Leini’, Torino, Italy

References