Using the modulating ion current gauge for total pressure measurements below 10−10 Torr without errors caused by electron-stimulated ion desorption
Source abstract
Abstract
Total pressure measurements below 10−10 Torr, in an active residual gas, were made using a special hot-cathode ionization gauge, without errors caused by electron-stimulated ion desorption. The gauge can separate the two ion signals, i.e., the signal created from gas phase molecules and that from electron-stimulated ion desorption, by using a new modulation and amplification system. In order to verify this separation effect, experiments were made with oxygen and nitrogen introduction in the 10−10 Torr range.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Analytical Chemistry and Sensors · Scientific Measurement and Uncertainty Evaluation · Mass Spectrometry Techniques and Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Sakuran-no-seibo Gakuin High School, Ohgida, Noda-machi, Fukushima-shi, 960, Japan
National Laboratory for High Energy Physics, Oho-machi, Tsukuba-gun, Ibaraki-ken, 305, Japan
References
Reference metadata was not supplied for this record.