journal article·1991-05-01

Relative sensitivity factors and useful yields for a microfocused gallium ion beam and time-of-flight secondary ion mass spectrometer

J. Bennett, D. Simons

Publisher page

Abstract

Relative sensitivity factors for several elements in micrometer size glass particles, bulk glasses and glass thin films have been determined for a Ga+ liquid metal ion gun combined with a time-of-flight secondary ion mass spectrometer. The reproducibility of the values within and among glasses of various compositions is better than 20%. Reproducibility is improved by sputter cleaning the sample prior to data collection. Relative sensitivity factors have also been obtained from bulk glasses using pulsed, low-energy electrons to neutralize charge buildup on the sample. Useful yields for Mg and Ca in a thin glass film were calculated to be 3.4×10−4 and 8.0×10−4, respectively.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Ion-surface interactions and analysis · Integrated Circuits and Semiconductor Failure Analysis · X-ray Spectroscopy and Fluorescence Analysis

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Surface and Microanalysis Science Division, National Institute of Standards and Technology, Gaithersburg, Maryland 20899

References

Reference metadata was not supplied for this record.