journal article·1996-11-01

Reduction of outgassing rate from residual gas analyzers for extreme high vacuum measurements

Shu Watanabe, Masakazu Aono, Shigeki Kato

Publisher page

Abstract

One problem associated with partial pressure measurement in an extreme high vacuum region is outgassing from the residual gas analyzer (RGA) itself. In order to reduce the outgassing rate from the RGA, the temperature of the filament should be lowered, and the heat generated around the ion source of the RGA should be dissipated in the atmosphere. We modified a RGA and reduced its outgassing rate to 1.5×10−12 Pa m3/s. We also determined the partial outgassing rate from the ion source and chamber quantitatively and found that hydrogen constituted more than 99% of the outgassing from the stainless steel.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Advanced Sensor Technologies Research · Advanced Electrical Measurement Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

The Institute of Physical and Chemical Research (RIKEN), Wako-shi, Saitama 351-01, Japan

National Laboratory for High Energy Physics (KEK), Tsukuba, Ibaraki 350, Japan

References

Reference metadata was not supplied for this record.