journal article·1984-10-01Open access

Design of chromel–alumel thermocouple gauges

Celso M. Aldao, Daniel G. Löffler

Abstract

The operation principle of thermocouple gauges is well known [Ref. 1: D. R. Denison, Methods of Experimental Physics, (Academic, New York, 1979)] but values of the physical parameters needed for the design of these instruments are not readily available in the literature. We have measured the electrical resistivity, thermal conductivity, and emissivity for chromel and alumel wires as functions of temperature. These functions can be used to design thermocouple gauges presenting maximum sensitivity in a particular pressure range or to modify the sensitivity of commercial instruments for a specific application.
Source: crossref · Quality: baseline