Properties of pin hole orifices as transfer leak artifacts in the 10−11 to 10−6 mol/s range
Source abstract
Abstract
A simple leak artifact has been developed using a single orifice only a few microns in diameter. The new single orifices have been calibrated over several orders of magnitude in leak rate by varying the gas pressure applied to one side of the element. The short term stability of the leak rate is limited primarily by the stability of the applied pressures. Long term stability appears to be limited by the cleanliness of the orifice and calibration system. The typical calibration curves obtained will be presented.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Advanced Sensor Technologies Research · Advanced MEMS and NEMS Technologies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Sandia National Laboratories, Albuquerque, New Mexico 87185