Abstract

This paper describes a new type of hot filament ionization gauge for measurement of neutral gas density in a high magnetic field. The notable feature of the gauge is its spherical symmetry. The ion current varies linearly with the pressure from 10−4 to 1 Pa for N2 and Ar. The sensitivity changes by no more than +20% in a magnetic field of up to 1.2 T. The main reasons for this small change in the sensitivity are as follows: (i) The probability of ionization of gas molecules is almost independent of a high magnetic field, because the spherical symmetry causes electrons to be emitted from the filament to the region in which the electric field is almost parallel to the magnetic field; (ii) the probability of ion collection is independent of a magnetic field.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Atomic and Molecular Physics · Mass Spectrometry Techniques and Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Naka Fusion Research Establishment, Japan Atomic Energy Research Institute, Naka-machi, Naka-gun, Ibaraki-ken 311-01, Japan