Clean ultrahigh vacuum system with single-structure diffusion pumps
Source abstract
Abstract
A practical example of the use of single-structure diffusion pumping groups in an experiment requiring hydrocarbon-free, ultrahigh vacuum is presented. With the use of proper vacuum procedures and careful foreline design and operation, but with no exotic hardware or liquid nitrogen, we have demonstrated that such a pump could provide vacuum that was ‘‘clean,’’ i.e., which had no residual pump-oil contamination to a partial pressure below 10−8 mbar. Their use is thus compatible, at least in peripheral vacuum chambers, with standard ‘‘UHV-surface science’’ processes. We discuss our vacuum system in detail, and present a review of the literature on the production of clean vacua with diffusion pumps.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Quantum, superfluid, helium dynamics · Heat Transfer and Boiling Studies · Phase Equilibria and Thermodynamics
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Laboratory for Atomic and Molecular Research and the Physics Department, University of Missouri–Rolla, Rolla, Missouri 65401
References
Reference metadata was not supplied for this record.