Portable total reflection x-ray fluorescence analysis in the identification of unknown laboratory hazards
Source abstract
Abstract
In this study, a portable total reflection x-ray fluorescence (TXRF) spectrometer was used to analyze unknown laboratory hazards that precipitated on exterior surfaces of cooling pipes and fume hood pipes in chemical laboratories. With the aim to examine the accuracy of TXRF analysis for the determination of elemental composition, analytical results were compared with those of wavelength-dispersive x-ray fluorescence spectrometry, scanning electron microscope and energy-dispersive x-ray spectrometry, energy-dispersive x-ray fluorescence spectrometry, inductively coupled plasma atomic emission spectrometry, x-ray diffraction spectrometry (XRD), and x-ray photoelectron spectroscopy (XPS). Detailed comparison of data confirmed that the TXRF method itself was not sufficient to determine all the elements (Z > 11) contained in the samples. In addition, results suggest that XRD should be combined with XPS in order to accurately determine compound composition. This study demonstrates that at least two analytical methods should be used in order to analyze the composition of unknown real samples.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: X-ray Spectroscopy and Fluorescence Analysis · Electron and X-Ray Spectroscopy Techniques · Non-Destructive Testing Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Materials Science and Engineering, Kyoto University , Yoshida-Honmachi, Sakyo-ku, Kyoto 606-8501, Japan
Institute for Chemical Research, Kyoto University , Gokasho, Uji, Kyoto 611-0011, Japan
Kyoto Prefectural Technology Center for Small and Medium Enterprises , Chudojiminami machi, Shimogyo-ku, Kyoto 600-8813, Japan
References (26)
Show source-provided references
1. Rev. Sci. Instrum. · 1971 · 42 · 106910.1063/1.1685282 ↗
2. 1997
3. Spectrochim. Acta, Part B · 1997 · 52 · 90110.1016/S0584-8547(96)01674-6 ↗
4. Spectrochim. Acta, Part B · 2006 · 61 · 112910.1016/j.sab.2006.08.010 ↗
5. Anal. Chem. · 2002 · 74 · 453210.1021/ac025720y ↗
6. J. Vac. Sci. Technol. A · 1996 · 14 · 192410.1116/1.580361 ↗
7. Appl. Surf. Sci. · 2011 · 257 · 992510.1016/j.apsusc.2011.06.109 ↗
8. Spectrochim. Acta, Part B · 2008 · 63 · 135510.1016/j.sab.2008.10.002 ↗
9. Spectrochim. Acta, Part B · 2003 · 58 · 212910.1016/S0584-8547(03)00197-6 ↗
10. Spectrochim. Acta, Part B · 2006 · 61 · 114110.1016/j.sab.2006.06.007 ↗
11. Spectrochim. Acta, Part B · 2001 · 56 · 212710.1016/S0584-8547(01)00284-1 ↗
12. Anal. Chem. · 2007 · 79 · 259310.1021/ac062279t ↗
13. Analyst [London] · 2010 · 135 · 190910.1039/c0an00009d ↗
14. Adv. X-Ray Chem. Anal., Jpn. · 2010 · 41 · 29
15. Anal. Sci. · 2013 · 29 · 79310.2116/analsci.29.793 ↗
16. 2012
17. 2009
18. Adv. X-Ray Chem. Anal., Jpn. · 2013 · 44 · 81
19. Adv. X-Ray. Chem. Anal., Jpn. · 2006 · 37 · 281
20. Tetsu-to-Hagane · 2011 · 97 · 8110.2355/tetsutohagane.97.81 ↗
21. Mater. Des. · 2011 · 32 · 280310.1016/j.matdes.2010.12.056 ↗
22. Aluminium: Physical properties, characteristics and alloys
23. 2012 · 23
24. 2013 · 14
25. Met. Sci. Heat Treat. · 1964 · 6 · 22310.1007/BF00650661 ↗
26. 2003 · 437