journal article·2011-07-01

Precision vacuum pumping speed measurement using sonic nozzles

Wan-Sup Cheung, Jin-Hyun Shin, Kyung-Am Park, Jong-Yeon Lim

Publisher page

Abstract

Constant volume flow meter systems were developed for measuring pumping speeds of dry vacuum pumps. Their capability to measure throughput in the low vacuum region was exploited to calibrate 0.03, 0.2, and 1.6 mm throat diameter sonic nozzles. It was demonstrated that the calibrated sonic nozzles enabled the precise measurement of the pumping speed of a dry vacuum pump under test as the constant volume flow meter systems. To examine the measurement capability of both approaches considered in this study, the expanded measurement uncertainty, as well as the detailed measurement uncertainty budget, is presented. The proposed pumping speed measurement method has potential application for a new “in situ and portable” pumping speed tester or analyzer for the Korean semiconductor and flat display process industry.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Advanced Sensor Technologies Research · Flow Measurement and Analysis

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Korea Research Institute of Standards and Science (KRISS) , 1 Doryong-Dong, Yuseong, Daejeon, South Korea

References (6)

Show source-provided references
1. International Organization for Standardization · 2005
2. Key Eng. Mater. · 2005 · 227–279 · 100010.4028/www.scientific.net/KEM.277-279.1000 ↗
3. International Organization for Standardization · 2007
4. J. Vac. Sci. Technol. A · 1987 · 5 · 255210.1116/1.574420 ↗
5. 1992
6. International Organization for Standardization · 2005