Optimizing charge neutralization for a magnetic sector SIMS instrument in negative mode
Source abstract
Abstract
Successful self-adjusted charge compensation was demonstrated for a CAMECA magnetic-sector secondary ion mass spectrometer applied in negative mode. Operation with the normal-incidence electron gun (NEG) potential positively biased relative to a sample potential enables substantial broadening of the Cs primary-ion-current density range available for analysis of insulators. The decrease of the negative NEG potential by 30 V allows the highest value of primary current density used for the analysis of a silica sample to increase by a factor of more than 6. By applying the improved charge neutralization technique, accurate Na depth profiles for SiO2 samples were obtained within detection limits of ∼3 × 1015 atoms/cm3.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Ion-surface interactions and analysis · Electron and X-Ray Spectroscopy Techniques · Analytical chemistry methods development
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Corning Incorporated , SP-FR-01-01, Corning, New York 14831
References (10)
Show source-provided references
1. Surf. Interface Anal. · 1990 · 16 · 910.1002/sia.740160105 ↗
2. Appl. Phys. Lett. · 1978 · 33 · 19310.1063/1.90271 ↗
3. Phys. Rev. B · 1974 · 10 · 263210.1103/PhysRevB.10.2632 ↗
4. Appl. Phys. Lett. · 1979 · 34 · 44410.1063/1.90828 ↗
5. J. Vac. Sci. Technol. A · 2000 · 18 · 27310.1116/1.582146 ↗
6. Benninghoven · 1994 · 864
7. Benninghoven · 1997 · 495
8. Benninghoven · 1989 · 655
9. Benninghoven · 2000 · 593
10. 1989 · App