Microfabrication of integrated scanning tunneling microscope
Source abstract
Abstract
An integrated circuit-compatible process has been demonstrated for microfabricating arrays of scanning tunneling microscopes on silicon wafers. Each microfabricated scanning tunneling microscope (STM) incorporates a novel three-dimensional thin-film piezoelectric actuator which is capable of scanning the STM tip laterally over the sample surface and following surface topography. The actuator is in the form of a cantilever bimorph constructed from alternating layers of multiple metallic electrodes, dielectric films, and piezoelectric zinc oxide films. Each device is designed to scan a region 0.1×1 μm with low voltage drive, with a maximum vertical deflection of 15 μm, which simplifies sample approach. The dimensions of the device are 1000×200×8 μm. Operation of the device has been demonstrated by imaging graphite with atomic resolution. The piezoelectric actuator in this device is itself a significant contribution to the field of microactuators, with many non-STM applications.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Force Microscopy Techniques and Applications · Surface and Thin Film Phenomena · Mechanical and Optical Resonators
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Applied Physics, Stanford University, Stanford, California 94305
References