Safety interlock and remote computer monitoring for a diffusion-pumped vacuum system
Source abstract
Abstract
A safety interlock system for a diffusion-pumped ultra-high vacuum chamber is presented. It controls and monitors vacuum system valves, diffusion pumps, liquid nitrogen filling, chilled water flow, and pressures in the chamber. The interlock system is interfaced with a phone dialer and a personal computer to provide remote monitoring and diagnostics of the vacuum system. This interlock prevents many operational errors and problems due to the power and liquid nitrogen failure.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Cardiac pacing and defibrillation studies · Microfluidic and Capillary Electrophoresis Applications · Molten salt chemistry and electrochemical processes
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Environmental Health and Toxicology, Wadsworth Center, State University of New York, Albany, New York 12201-0509
References
Reference metadata was not supplied for this record.