Use of successive expansion technique to determine volume ratios of ≊5000 for vacuum gauge calibration
Source abstract
Abstract
The method of successive expansion first used by Elliott and Clapham [National Physical Laboratory (U.K.) Report No. MOM28, 1978] for measuring volume ratios of ∼100 has been shown to be feasible for measuring volume ratios of ∼2830 with uncertainties of ±0.15 % while using a capacitance manometer with a lower pressure range for measuring the final pressures. The volume ratios thus determined have enabled the authors to reduce the systematic errors associated with the pressures generated by the series expansion system in their laboratory. Further, an attempt has been made to quantify the error in the measurement of volumes of chambers consisting of one or more metal diaphragm valves. In order to completely eliminate errors on this account an alternative calibration procedure has been outlined. The possibility of extending the measurement to ratios of ≊5000 has also been discussed.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Advanced Sensor Technologies Research · Advanced Electrical Measurement Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
National Physical Laboratory, New Delhi-110012, India
References