In situ bend fixture for deformation and fracture studies in the environmental scanning electron microscope
Source abstract
Abstract
A computer controlled loading fixture has been designed to allow in situ observation of fracture processes during bending deformation of metal/ceramic microlaminates in an electroscan environmental scanning electron microscope (ESEM). The stage has the capability of accommodating either three- or four-point bending experiments. A unique design feature of the stage is that the specimen surface remains at a fixed distance from the secondary electron detector and, hence, in focus during bending. The sample rests on the fulcrum which remains in a fixed position while the restraints that grip the ends of the sample descend on a ball slide. The system is controlled by a Macintosh IIci computer and National Instruments NB-MIO-16L-9 data acquisition card. National Instruments LabVIEW(R)2 software is used to control the stage displacement and to record the load cell and transducer outputs. The operation of this instrumentation in the ESEM is illustrated by the study of fracture processes in ceramic and ceramic/metal microlaminate films deposited on ductile metallic substrates.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Electron and X-Ray Spectroscopy Techniques · Corrosion Behavior and Inhibition · Environmental Impact and Sustainability
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References