I n s i t u cleaning of nickel field-ion surfaces by neon ion bombardment
Source abstract
Abstract
Neon ion bombardment followed by field evaporation is shown to be a simple, effective procedure for cleaning the surface of nickel field-ion samples. The neon ions are generated by reversing the tip polarity and establishing a field emission current in the presence of the neon imaging gas. The procedure removes surface impurities which are resistant to field desorption at the evaporation field of nickel. The experimental conditions required to clean a nickel surface are presented.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Ion-surface interactions and analysis · Integrated Circuits and Semiconductor Failure Analysis · Advanced Materials Characterization Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Sandia National Laboratories, Albuquerque, New Mexico 87185