Visualization of a buried organic interface by imaging time-of-flight secondary ion mass spectrometry and scanning Auger microprobe of an ion-beam crater edge
Source abstract
Abstract
A method has been developed for chemically imaging a buried organic interface of thickness similar to that of a biological conditioning film. Using an ion beam to form a sputter crater, elemental analysis of the edge of the crater is performed using Auger electron spectroscopy and chemical imaging of the edge is performed by time-of-flight secondary ion mass spectrometry. The crater edge that is formed due to the shape of the ion beam and geometry of the spectrometer has the ability to magnify the interfaces by 1000×. Samples mimicking titanium bone implant protheses were analyzed to demonstrate the feasibility of the method.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Ion-surface interactions and analysis · Integrated Circuits and Semiconductor Failure Analysis · Electron and X-Ray Spectroscopy Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Natural Sciences and Mathematics Complex, Department of Chemistry, SUNY at Buffalo, Buffalo, New York 14260-3000
Department of Biomaterials, SUNY at Buffalo, Buffalo, New York 14260-3000
Natural Sciences and Mathematics Complex, Departments of Chemistry and Biomaterials, SUNY at Buffalo, Buffalo, New York 14260-3000
Physical Electronics, Eden Prairie, Minnesota 55344
References
Reference metadata was not supplied for this record.