journal article·1997-05-01

Approach for a three-dimensional on-chip quantification by secondary-ion mass spectrometry analysis

Hubert Gnaser

Publisher page

Abstract

An attempt to perform a three-dimensional (3D) quantitative evaluation of the complete sample volume sputter removed in a secondary-ion mass spectrometry analysis is described. Using a focused (∼3 μm) 7 keV Cs+ primary-ion beam and detecting sputtered MCs+ ions (M stands for any sample element), a vertical Si–metal–oxide–semiconductor transistor structure (with a lateral size of ∼40 μm and a total vertical dimension of ∼3 μm) was analyzed by image depth profiling. Elemental concentrations of these micrographs were derived from relative sensitivity factors and the pertinent depth information in the 3D sample (data) volume could be estimated by utilizing these factors and an approximation of the atomic densities. The specimen’s surface topography both in the pristine state and upon completion of the sputter depth profile was recorded by two-dimensional stylus profilometry. By these means, a reconstruction of the sample volume eroded during the analysis appears feasible at a lateral resolution of a few μm and an in-depth resolution of about 10 nm.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Ion-surface interactions and analysis · Integrated Circuits and Semiconductor Failure Analysis · Electron and X-Ray Spectroscopy Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Institut für Oberflächen hen-und Schichtanalytik and Fachbereich Physik, Universität Kaiserslautern, D-67663 Kaiserslautern, Germany

References

Reference metadata was not supplied for this record.