journal article·1992-07-01

Contamination control in the design and manufacture of gas flow components

John Sullivan, Skip Schaffer, Steve King, Dennis Manos, H. F. Dylla

Publisher page

Abstract

To assure the cleanliness of gas flow components such as mass flow meters and controllers, there must be an initial concentrated effort on the design for cleanliness. The design effort must include all elements, such as materials, sensor, and control valve technologies. The design must then be implemented by an appropriate manufacturing philosophy, facility, and etiquette. We describe in detail the design considerations, cleaning procedures, and quality assurance measurement tools and methods used in the process. This methodology is compared with alternative techniques, and is shown to produce gas flow components that indeed verify the conceptual ideas. Experimental data show the results of surface analysis (profilometry, x-ray photoelectron spectroscopy, scanning electron microscopy), particulate measurements using both laser particle and condensation nucleus, and moisture analysis (vibrating quartz crystal) instrumentation.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Advanced Sensor Technologies Research · Flow Measurement and Analysis

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

MKS Instruments, Inc., Andover, Massachusetts 01810

Princeton Scientific Consultants, Princeton, New Jersey 08543