A scanning tunneling microscope for ultrahigh vacuum atom–surface interaction studies
Source abstract
Abstract
A compact, mechanically stable and thermally compensated tunneling microscope has been designed for use in an ultrahigh vacuum (UHV) surface characterization and analysis system. Very small concentric tubular scanning elements provide mechanical rigidity, thermal stability, and minimal cross coupling of the X, Y, and Z motions. The compact load-locking design provides full compatibility with the rest of the UHV analysis system without requiring that a disproportionately large fraction of the vacuum system be devoted to the microscope. Reported here is the design of the microscope and load-lock system, and the results of tests using the microscope in air to image the surfaces of graphite, gold, and platinum.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Surface and Thin Film Phenomena · Advanced Materials Characterization Techniques · Force Microscopy Techniques and Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Laboratory of Atomic and Solid State Physics, Clark Hall, Cornell University, Ithaca NY 14853-2501