X-ray gauge measures areal density variations as small as 0.1%
Source abstract
Abstract
A gauge measures areal density variations as small as 0.1% in low density plastic foams and other materials. It has a spatial resolution of 0.5 mm. This system is safe, portable, easy to use, and is readily automated. The 6-keV x-ray flux from a 55Fe source is collimated to a 0.5-mm-diameter beam and attenuated by the sample. Transmitted x rays are detected with a proportional gas detector and counted with a scaler-timer. After measuring the x-ray attenuation in two areas of the sample, the areal density variation is calculated using an expression derived from I/I0=exp(−upx), where I is the transmitted flux, I0 is the incident flux, u is the mass attenuation coefficient of the sample, p is the foam density, x is the thickness of the sample, and the product px=D is the areal density. Measurements sensitive to 1% and 0.1% areal density variations take 3 min and 4 h per point, respectively, in 0.05-g/cm3 plastics, 1 cm thick. Absolute areal density is also measured to ±3% using existing x-ray attenuation tables. However, better accuracy may be obtained by calibrating the system with known samples.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Advanced X-ray and CT Imaging · X-ray Spectroscopy and Fluorescence Analysis · Nuclear Physics and Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Materials Science and Technology Division, Los Alamos National Laboratory, Los Alamos, New Mexico, 87545