A low-cost high performance low-energy electron diffraction intensity measuring apparatus with video data acquisition rate

Zhi-yuan Zhang, Jia-ho Lu, Si-min Chai

Publisher page

Abstract

In order to get the low-energy electron diffraction (LEED) I–E curves of a time-varying surface structure, we developed a well-designed multichannel programmable data acquisition system. By using a 2 μs (ADC) chip and some other hardwares the video data acquisition rate is realized. Up to 20 beams could be acquired. The tracking of spots is accomplished by the computation of the spots’ position. The size of the sampling window is slectable in the range of (4×4) to (16×16) pixels according to the size of the spot. Considering the inertia of the camera tube and the LEED gun supply, a pause of one-half to several tens of frames may be assigned after the control of energy is setup. In this paper, I–E curves of diffraction spots of Si (111)-(7×7) structure, including that of individual fractional spots, are demonstrated. This system can also measure the intensity distribution pattern of diffraction, the spot intensity variation of time I–t curve, and may be used for reflection high-energy electron diffraction (RHEED) as well.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Electron and X-Ray Spectroscopy Techniques · Advanced Electron Microscopy Techniques and Applications · Advancements in Photolithography Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Radio and Electronics Department, Tsinghua University, Beijing, People’s Republic of China

References