journal article·1988-09-01

Automated system for measuring leak standards

John M. Ball

Publisher page

Abstract

An inexpensive, automatic measurement system for standard leaks has been developed and tested. The principal of operation is pressure rise in a measurement chamber held at a low pressure. This design minimizes temperature effects, considerably improving performance over what would be achievable using a similar technique at atmospheric pressure. Provision is made for determining working measurement volume by expansion from a calibrated reference volume incorporated in the piping. Demonstrated operating range is ∼10−8–10−4 atm cm3/s (10−13–10−9 mol/s) regardless of the gas being used. The sensing element is an industrial grade capacitance diaphragm manometer. A very inexpensive home computer performs all control and calculations. The accuracy and performance of this system have been verified through comparison to other leak calibrators.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Analytical Chemistry and Sensors · Sensor Technology and Measurement Systems

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

U.S. Army Primary Standards Laboratory, Redstone Arsenal, Alabama 35898-5400

References