THE LOOP TECHNIQUE FOR STRAIN GAGE ROSETTE SIGNAL CONDITIONING
Source abstract
Abstract
Background The classical approach for signal conditioning individual resistance strain gages is to connect each installed gage with three lead wires to a set of three fixed completion resistors, all arranged electrically in a Wheatstone bridge circuit. An alternative approach for strain gage signal conditioning, sometimes called the Anderson loop, was developed at NASA's Dryden Flight Research Center. This technique offers some significant advantages when used with the strain gage rosettes employed in experimental stress analysis. Invented to eliminate the measurement uncertainties due to random electrical impedance variations in lead wires and connections found in harsh test environments, NASA's loop circuit approach delivers twice the equivalent bridge output level, and simplifies wiring, temperature compensation and data analysis as well. Circuitry The enabling technology for this technique is the dual-differential subtractor measurement circuit function defined in Fig. 1. As illustrated in eq. 1, a subtractor's output is the difference between two independent floating differential inputs.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Structural Health Monitoring Techniques · Advanced Electrical Measurement Techniques · Sensor Technology and Measurement Systems
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (7)
Show source-provided references
1. 1. Karl F. Anderson, "The Constant Current Loop: A New Paradigm for Resistance Signal Conditioning," NASA TM-104260, 1992 .
2. 10.1109/19.67671110.1109/19.676711 ↗
3. 3. Karl F. Anderson, Constant Current Loop Impedance Measuring System That Is Immune to the Effects of Parasitic Impedances, U.S. Patent No. 5,731,469, 1994 .
4. 4. Karl F. Anderson, "A Conversion of Wheatstone Bridge to Current-Loop Signal Conditioning for Strain Gages," NASA TM- 104309, 1995 .
5. 10.1109/5289.65827010.1109/5289.658270 ↗
6. 6. Karl F. Anderson, Continuous Measurement of Both Thermoelectric and Impedance Based Signals Using Either AC or DC Excitation , Measurement Science Conference, 1997 .
7. 7. Trig-Tek, Inc. Model 5905 Rosette Current Loop Conditioner Data Sheet, http://www.trig-tek.com.