Micromolding fabrication of biocompatible dry micro-pyramid array electrodes for wearable biopotential monitoring
Source abstract
Abstract
Abstract Thin dry electrodes are promising components in wearable healthcare devices. Assessing the condition of the human body by monitoring biopotentials facilitates the early diagnosis of diseases as well as their prevention, treatment, and therapy. Existing clinical-use electrodes have limited wearable-device usage because they use gels, require many preparation steps, and can be uncomfortable to wear. Dry electrodes can improve these issues and have demonstrated performance on par with gel-based electrodes, providing advantages in mobile and wearable applications. However, the materials and fabrication methods used are not yet at the level of disposable gel electrodes for low-cost mass manufacturing and wide adoption. Here, a low-cost manufacturing process for thin dry electrodes with a conductive micro-pyramidal array (MPA) is presented for large-scale on-skin wearable applications. The electrode is fabricated using micromolding techniques in conjunction with solution processes in order to guarantee ease of fabrication, high device yield, and the possibility of mass production compatible with current semiconductor production processes. Fabricated using a conductive paste and an epoxy resin that are both biocompatible, the developed MPA electrode operates in a conformal, non-invasive manner, with low skin irritation, which ensures improved comfort for brief or extended use. The operation of the developed electrode was examined by analyzing electrode-skin-electrode impedance, electroencephalography, electrocardiography, and electromyography signals and comparing them with those measured simultaneously using gel electrodes.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Advanced Sensor and Energy Harvesting Materials · Neuroscience and Neural Engineering · Conducting polymers and applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (38)
Show source-provided references
1. Trends in epidermal stretchable electronics for noninvasive long-term healthcare applications10.5875/ausmt.v7i2.1336 ↗
2. Skin conformal polymer electrodes for clinical ECG and EEG recordings10.1002/adhm.201700994 ↗
3. Wearable and implantable devices for cardiovascular healthcare: from monitoring to therapy based on flexible and stretchable electronics10.1002/adfm.201808247 ↗
4. Flexible inkjet-printed multielectrode arrays for neuromuscular cartography10.1002/adhm.201600108 ↗
5. Focal stimulation of the sheep motor cortex with a chronically implanted minimally invasive electrode array mounted on an endovascular stent10.1038/s41551-018-0321-z ↗
6. Ultraflexible organic amplifier with biocompatible gel electrodes10.1038/ncomms11425 ↗
7. Nanomaterial-enabled dry electrodes for electrophysiological sensing: a review10.1007/s11837-016-1818-0 ↗
8. Dry-contact and noncontact biopotential electrodes: methodological review10.1109/RBME.2010.2084078 ↗
9. Wireless soft scalp electronics and virtual reality system for motor imagery-based brain-machine interfaces10.1002/advs.202101129 ↗
10. Process optimization and characterization of silicon microneedles fabricated by wet etch technology10.1016/j.mejo.2005.04.044 ↗
11. Microfabricated microneedles: a novel approach to transdermal drug delivery10.1021/js980042+ ↗
12. Biodegradable polymer microneedles: fabrication, mechanics and transdermal drug delivery10.1016/j.jconrel.2005.02.002 ↗
13. Nano-layered microneedles for transcutaneous delivery of polymer nanoparticles and plasmid DNA10.1002/adma.201001525 ↗
14. Rapid fabrication of microneedles using magnetorheological drawing lithography10.1016/j.actbio.2017.10.030 ↗
15. A low contact impedance medical flexible electrode based on a pyramid array micro-structure10.3390/mi11010057 ↗
16. Single-step fabrication of computationally designed microneedles by continuous liquid interface production10.1371/journal.pone.0162518 ↗
17. Novel flexible wearable sensor materials and signal processing for vital sign and human activity monitoring10.3390/s17071622 ↗
18. Scalp electrode impedance and EEG data quality10.1016/S1388-2457(00)00533-2 ↗
19. Fabrication of dissolving polymer microneedles for controlled drug encapsulation and delivery: bubble and pedestal microneedle designs10.1002/jps.22140 ↗
20. Microneedles for drug and vaccine delivery10.1016/j.addr.2012.04.005 ↗
21. Updated CDC recommendations for the management of hepatitis B virus-infected health-care providers and students
22. Fully organic compliant dry electrodes self-adhesive to skin for long-term motion-robust epidermal biopotential monitoring10.1038/s41467-020-18503-8 ↗
23. Navaraj · 202010.1016/B978-0-08-102407-2.00006-0 ↗
24. Conductive and stretchable adhesive electronics with miniaturized octopus-like suckers against dry/wet skin for biosignal monitoring10.1002/adfm.201805224 ↗
25. Wearable silver nanowire dry electrodes for electrophysiological sensing10.1039/C4RA15101A ↗
26. Ferromagnetic, folded electrode composite as a soft interface to the skin for long-term electrophysiological recording10.1002/adfm.201603146 ↗
27. Self-adhesive epidermal carbon nanotube electronics for tether-free long-term continuous recording of biosignals10.1038/srep06074 ↗
28. Soft, Comfortable polymer dry electrodes for high quality ECG and EEG recording10.3390/s141223758 ↗
29. Gas phase deposition of trichloro(1H,1H,2H,2H-perfluorooctyl)silane on silicon dioxide, by XPS10.1116/11.20071103 ↗
30. Organometallic chemistry on silicon and germanium surfaces10.1021/cr000064s ↗
31. Residual-layer-free printing by selective filling of self-assembled monolayer-treated mold10.1116/1.3596558 ↗
32. 10993–11: 2006 biological evaluation of medical devices—part 11: tests for systemic toxicity
33. 10993–10: 2010 biological evaluation of medical devices—part 10: tests for irritation and skin sensitization
34. 10993–6: 2016 biological evaluation of medical devices—part 6: tests for local effects after implantation
35. 10993–5: 2009 biological evaluation of medical devices—part 5: tests for in vitro cytotoxicity
36. ASTM F619-03: 2008 standard practice for assessment of hemolytic properties of materials
37. Matrix-independent highly conductive composites for electrodes and interconnects in stretchable electronics10.1021/acsami.8b21836 ↗
38. Keenan · 2011 · 1602