Design and detection method of a four electrodes cross-conductive sensor for fluid conductivity measurement
Source abstract
Abstract
In this paper, two types of cross-conductive structures have been designed and simulated in COMSOL MULTIPHYSICS 5.6 software. The effectiveness of the structure has been evaluated by varying (i) thickness (t) (ii) electrode material (iii) gap between the electrodes and (iv) the effective area of the structure. The structure has been validated by various standard conducting materials of known conductance. Film of proper thickness has been deposited in the various structure during the simulation. The conductivity of conducting film based on cross-conductance per unit length is then evaluated by extension of Thompson and Lampard Theorem (TL). The simulation result and theoretically calculated value are found in good agreement. The shielding of the cross structure reduces the error. For effective measurement, the electrode conductivity must be very high as compared to film material. A cross structure has been fabricated and the conductivity of the commercial and groundwater samples are measured.
Source: expertdata · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Analytical Chemistry and Sensors · Electrochemical Analysis and Applications · Gas Sensing Nanomaterials and Sensors
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (30)
Show source-provided references
1. G.H. Rayner, NPL calculable capacitor. IEEE Trans. Instrum. Meas. 21(4), 361–365 (1972)10.1109/TIM.1972.4314044 ↗
2. A.M. Thompson, D.G. Lampard, A new theorem in electrostatics and its application to calculable standards of capacitance. Nature 177(4515), 888–888 (1956)10.1038/177888a0 ↗
3. Yu. Jing, Yu. Hang, D. Li, Design and characteristic analysis of cross-capacitance fuel-level sensor. Sensors 18(11), 3984 (2018)10.3390/s18113984 ↗
4. R.D. Lee, H.J. Kim, Y.P. Semenov, Precise measurement of the dielectric constants of liquids using the principle of cross capacitance. IEEE Transact. Instrum. Measurement 50(2), 298–30 (2001)10.1109/19.918126 ↗
5. Z.H. Zargar, T. Islam, A novel cross-capacitive sensor for noncontact microdroplet detection. IEEE Transactions on Industrial Electronics 66(6), 4759–4766 (2018)10.1109/TIE.2018.2863205 ↗
6. Z.H. Zargar, T. Islam, A thin film porous alumina-based cross-capacitive humidity sensor. IEEE Transact. Instrumentation Measurement 69(5), 2269–2276 (2019)10.1109/TIM.2019.2921438 ↗
7. S. Afzal Ali, M. Rahat Mahboob, T. Islam, A simple setup for temperature study of Humidity Sensors, 2nd International Conference on Emerging Frontiers in Electrical and Electronic Technologies (ICEFEET) IEEE, 1–5 (2022).10.1109/ICEFEET51821.2022.9847881 ↗
8. T. Islam, M. Yousuf, M. Nauman, A highly precise cross-capacitive sensor for metal debris detection in insulating oil. Rev. Sci. Instrum. 91(2), 025005 (2020)10.1063/1.5139925 ↗
9. O. Rahman, T. Islam, N. Khera, S.A. Khan, A novel application of the cross-capacitive sensor in real-time condition monitoring of transformer oil. IEEE Transact. Instrum. Measurement 70, 1–12 (2021)
10. T. Islam, A. Salamat, S. Kumar Singh, M. Rehman, A cross-conductive sensor to measure bottled water quality, IEEE International Instrumentation and Measurement Technology Conference (I2MTC) IEEE, 1–6 (2021).10.1109/I2MTC50364.2021.9459929 ↗
11. Z. H. Zargar, T. Islam, K. J. Akram, Design and development of a non-contact cross-capacitive micro droplet detector, IEEE Sens., 1–4 (2018).10.1109/ICSENS.2018.8589702 ↗
12. T. Islam, O.P. Maurya, A.U. Khan, Design and fabrication of fringing field capacitive sensor for non-contact liquid level measurement. IEEE Sens. J. 21(21), 24812–24819 (2021)10.1109/JSEN.2021.3112848 ↗
13. M. Rehman, V.G.K. Murti, An extension and application of the Thompson-Lampard theorem. Proc. IEEE 69(11), 1512–1514 (1981)10.1109/PROC.1981.12190 ↗
14. M. Rehman, V.G.K. Murti, M.R. Gupta, A novel precision type liquid conductivity measuring system. J. Phys. E: Sci. Instrum. 18(11), 902 (1985)10.1088/0022-3735/18/11/004 ↗
15. M. N. Anas, A. N. Norali, M. Rehman, Development of a low-cost precision blood resistivity measurement device based on Thompson Lampard Theorem, International Conference on Biomedical Engineering (ICoBE) IEEE, 594–597 (2012).10.1109/ICoBE.2012.6178986 ↗
16. T. Islam, A. Salamat, S.K. Singh, M. Rehman, A direct AC cross conductive sensor for milk quality measurement. IEEE Trans. Instrum. Meas. 71, 1–8 (2022)
17. T. Islam, Z.H. Zargar, M. Rehman, A novel humidity sensor based on the extension of Thompson and Lampard theorem. IEEE Transact. Electron Devices 62(12), 4237–4241 (2015)10.1109/TED.2015.2492779 ↗
18. U. Singh, K. J. Akram, T. Islam, G. R. Biswal and P. K. Parida, Cross-Conductive Sensor for Humidity Measurement in Gas for Gas Insulated Switchgears Application, IEEE Second International Conference on Control, Measurement and Instrumentation (CMI) IEEE, 1–5 (2021).10.1109/CMI50323.2021.9362922 ↗
19. K. Xu, Y. Chen, T.A. Okhai, L.W. Snyman, Micro optical sensors based on avalanching silicon light-emitting devices monolithically integrated on chips. Optical Mater. Express 9(10), 3985–3997 (2019)10.1364/OME.9.003985 ↗
20. Z. Zargar, T. Islam, Design and modeling of cross capacitor based on Thompson and Lampard Theorem, National Conference on Recent Developments in Electronics (NCRDE) (2017)
21. J.-Y. Kim, D. Sohn, E.-R. Kim, Polymer-based multi-layer conductive electrode film for plastic LCD applications. Appl. Phys. A 72(6), 699–704 (2001)10.1007/s003390100853 ↗
22. S. Felix, A novel CuO–N-doped graphene nanocomposite-based hybrid electrode for the electrochemical detection of glucose. Appl. Phys. A 123(10), 1–9 (2017)10.1007/s00339-017-1217-6 ↗
23. O.A. Al-Hartomy, Highly conductive polyaniline/graphene nano-platelet composite sensor towards detection of toluene and benzene gases. Appl. Phys. A 125(1), 1–9 (2019)10.1007/s00339-018-2317-7 ↗
24. S. Ramanathan, Elastomeric polydimethylsiloxane polymer on conductive interdigitated electrode for analyzing skin hydration dynamics. Appl. Phys. A 126(9), 1–11 (2020)10.1007/s00339-020-03933-4 ↗
25. T. Sarkar, N. Mukherjee, J. Das, Glucose oxidase immobilized macro porous silicon based conductive glucose sensor. Appl. Phys. A 128(4), 1–10 (2022)10.1007/s00339-022-05453-9 ↗
26. C.L. Chen, Y.C. King, TiN metal gate electrode thickness effect on BTI and dielectric breakdown in HfSiON-based MOSFETs. IEEE Transact Electron Devices 58(11), 3736–3742 (2011)10.1109/TED.2011.2163819 ↗
27. COMSOL Multiphysics, Introduction to COMSOL Multiphysics, accessed 2018 (1998): 32 https://www.comsol.com. Accessed 01 April 2023
28. A. R. Mohd Syaifudin, S. C. Mukhopadhyay, P. L. Yu, Electromagnetic field computation using COMSOL multiphysics to evaluate the performance of novel interdigital sensors, IEEE 1–4 (2010)10.1109/AEMC.2009.5430661 ↗
29. Conductivity of water, https://byjus.com/physics/conductivity-of-water/. Accessed 01 April 2023
30. MatWeb Material Property Data, https://www.matweb.com/. Accessed 01 April 2023