The aim of this study was to assess the release of metal elements from a Ti6Al4V modular endoprosthesis for mandibular reconstruction. Ten monkeys were included, seven of the animals had an endoprosthesis inserted and three served as controls. Mucosa, regional lymph nodes and distant organs, were assessed after the implant had been in place for 12 months, using light (LM) and transmission electron microscopy (TEM) and inductively coupled plasma optical emission spectroscopy (ICP). Blood was also drawn from all animals for elemental analysis using ICP. LM and TEM evaluation showed no detectable metal particles in the mucosa surrounding the endoprosthesis, nor in the regional lymph nodes and distant organs. Blood analysis revealed that titanium and vanadium were detectable in comparable amounts in the test (Ti: 1.63+/-0.54, Va: 0.42+/-0.08) as well as in the control group (Ti: 2.07+/-0.55, Va: 0.37+/-0.07). The amount of aluminum appeared to be higher in the control group (31.77+/-11.67) compared with the test group (20.41+/-9.13), but this difference was not statistically significant. ICP showed that no titanium, vanadium or aluminum was detectable in the mucosa surrounding the endoprosthesis. In addition, no titanium and vanadium were found in the lymph nodes and distant organs using ICP. On the other hand, using ICP, the aluminum content was found to be higher in the right regional lymph nodes and all examined distant organs as compared to the control group (lymph nodes: 11.55+/-22.15 vs. 0.36+/-0.61, lung: 6.24+/-11.28 vs. 1.40+/-2.15, liver: 1.66+/-0.99 vs. 0, kidney: 15.69+/-24.88 vs 0, spleen: 2.75+/-3.09 vs. 0.49+/-0.43). However, only for the kidney and liver the higher amount of aluminum was statistically significant. In conclusion, this study quantified the release of only aluminum in lymph nodes and distant organs, when using a modular endoprosthesis made of Ti-6Al-4V for mandibular reconstruction.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Bone Tissue Engineering Materials · Orthopaedic implants and arthroplasty · Dental Implant Techniques and Outcomes
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (40)
Show source-provided references
1. The biological response to titanium and titanium–aluminium–vanadium alloy particles. II. Long-term animal studies10.1016/0142-9612(86)90086-4 ↗
2. Release of elements from retrieved maxillofacial plates and screws10.1007/s10856-005-3575-4 ↗
3. Tissue response to titanium plates: a transmitted electron microscopic study10.1016/S0278-2391(97)90115-4 ↗
4. Tissue changes adjacent to titanium plates in patients10.1054/jcms.2002.0286 ↗
5. Dynamic study about metal release from titanium miniplates in maxillofacial surgery10.1054/ijom.2000.0039 ↗
6. Element release from titanium devices used in oral and maxillofacial surgery10.1016/S0142-9612(02)00424-6 ↗
7. Difference in metallic wear distribution released from commercially pure titanium compared with stainless steel plates10.1007/s00402-003-0614-9 ↗
8. A study of titanium release into body organs following the insertion of single threaded screw implants into the mandibles of sheep10.1111/j.1834-7819.2002.tb00331.x ↗
9. Dissemination of wear particles to the liver, spleen, and abdominal lymph nodes of patients with hip or knee replacement10.2106/00004623-200004000-00002 ↗
10. Titanium deposition in regional lymph nodes after insertion of titanium screw implants in maxillofacial region10.1016/S0901-5027(05)80045-1 ↗
11. Metal release from titanium fixtures during placement in the mandible: an experimental study
12. Metal ion release from titanium-based prosthetic segmental replacements of long bones in baboons: a long-term study10.1002/jor.1100010411 ↗
13. Femoral bone loss in total knee arthroplasty. A review
14. Metal carcinogenesis in total joint arthroplasty. Animal models10.1097/00003086-199608001-00023 ↗
15. Immune responses correlate with serum-metal in metal-on-metal hip arthroplasty10.1016/j.arth.2004.09.012 ↗
16. Metallosis after contemporary metal-on-metal total hip arthroplasty. Five to nine-year follow-up10.2106/JBJS.D.02916 ↗
17. Cipriano CA, Issack PS, Beksac B, Della Valle AG, Sculco TP, Salvati EA. Metallosis after metal-on-polyethylene total hip arthroplasty. Am J Orthop (Belle Mead, NJ) 2008;37(2):E18–25.
18. Revision total hip arthroplasty in hip joints with metallosis: a single-center experience with 31 cases10.1016/j.arth.2005.04.001 ↗
19. A histologic and biochemical comparison of the interface tissues in cementless and cemented hip prostheses10.1097/00003086-199302000-00023 ↗
20. Aluminum, Alzheimer’s disease and bone fragility10.3109/17453679708999016 ↗
21. Boyce BF, Eider HY, Fell SG, Nicholson WA, Smith GD, Dempster DW, et al. Quantitation and localisation of aluminum in human cancellous bone in renal osteodystrophy. Scan Electron Microsc 1981;(Pt 3):329–37.
22. Modular endoprosthesis for mandibular reconstruction: a preliminary animal study10.1016/j.ijom.2008.07.021 ↗
23. Application of inductively coupled plasma sector field mass spectrometry for elemental analysis of urine10.1016/S0946-672X(01)80010-9 ↗
24. Brain aluminum, magnesium and phosphorus contents of control and Alzheimer-diseased patients10.3233/JAD-2005-7402 ↗
25. Silicon reduces aluminum accumulation in rats: relevance to the aluminum hypothesis of Alzheimer disease10.1097/00002093-199806000-00005 ↗
26. Is exposure to aluminum a risk factor for the development of Alzheimer disease?–Yes10.1001/archneur.55.5.740 ↗
27. Aluminum exposure and risk of Alzheimer disease10.1001/archneur.55.5.742 ↗
28. Fixation and metal release from the Tifit femoral stem prosthesis. 5-Year follow-up of 64 cases10.3109/17453679808999049 ↗
29. Widespread dissemination of metal debris from implants10.1302/0301-620X.76B5.8083255 ↗
30. Ultrastructural analysis of metal particles released from stainless steel and titanium miniplate components in an animal model10.1016/S0278-2391(98)90915-6 ↗
31. A retrieval study on morphological and chemical changes of titanium osteosynthesis plates and adjacent tissues10.1016/j.jcms.2007.01.004 ↗
32. Titanium lines: a manifestation of metallosis and tissue response to titanium alloy megaprostheses at the knee10.1016/S0009-9260(05)81138-9 ↗
33. Rodrigues DC, Urban RM, Jacobs JJ, Gilbert JL. In vivo severe corrosion and hydrogen embrittlement of retrieved modular body titanium alloy hip-implants. J Biomed Mater Res 2008.10.1002/jbm.b.31171 ↗
34. Kop AM, Swarts E. Corrosion of a hip stem with a modular neck taper junction: a retrieval study of 16 cases. J Arthroplasty 2008.
35. Local and distant products from modularity
36. Evaluation of the biological response of wear debris10.1016/j.spinee.2004.07.014 ↗
37. Determination of metallic traces in kidneys, livers, lungs and spleens of rats with metallic implants after a long implantation time10.1007/s10856-007-3002-0 ↗
38. Evaluation of methods for trace-element determination with emphasis on their usability in the clinical routine laboratory10.1080/00365510601095281 ↗
39. A consensus paper on metal ions in metal-on-metal hip arthroplasties10.1016/j.arth.2004.09.009 ↗
40. Titanium identification and measurement by ICP-MS and ICP-OES in a human spleen fragment10.1093/clinchem/42.11.1875 ↗