The relative effects of radiation crosslinking and type of counterface on the wear resistance of ultrahigh-molecular-weight polyethylene
Source abstract
Abstract
The lifetime of total joint replacement prostheses utilizing ultrahigh-molecular-weight polyethylene (UHMWPE) components has historically been determined by their wear resistance. It has been discovered that radiation crosslinking of UHMWPE can substantially increase its wear resistance. However, it is also well recognized that there is a radiation-dose-dependent decrease in several important mechanical properties of UHMWPE, such as fracture toughness and resistance to fatigue crack propagation. In this study, the effect of radiation crosslinking (followed by remelting) on the morphology, tensile properties and wear resistance of UHMWPE was investigated. Wear tests were conducted against both the commonly used cobalt-chromium counterface polished to implant grade smoothness as well as a smoother ceramic (alumina) counterface. The results showed that 50kGy dose radiation crosslinking increased the wear resistance of UHMWPE against the cobalt-chromium counterface 7-fold, but the coupling of remelted, crosslinked UHMWPE against the smoother alumina counterface led to a 20-fold increase in wear resistance. This study shows that the use of an alumina counterface would circumvent the need to use a high radiation dose in crosslinking UHMWPE, associated with poor mechanical properties, without compromising wear resistance.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced materials and composites · Total Knee Arthroplasty Outcomes
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (34)
Show source-provided references
1. The problem is osteolysis
2. The origin of submicron polyethylene wear debris in total hip arthroplasty
3. Oxidation of orthopaedic UHMWPE10.1016/S0142-9612(01)00288-5 ↗
4. In vivo UHMWPE biodegradation of retrieved prosthesis10.1016/S0142-9612(98)00013-1 ↗
5. The Otto Aufranc Award. Impact of gamma sterilization on clinical performance of polyethylene in the hip10.1097/00003086-199510000-00004 ↗
6. A study of the combined effects of shelf ageing following irradiation in air and counterface roughness on the wear of UHMWPE
7. Impact of gamma sterilization on clinical performance of polyethylene in the knee10.1016/S0883-5403(96)80026-X ↗
8. Quantification of third-body damage and its effect on UHMWPE wear with different types of femoral head10.1302/0301-620X.80B5.0800894 ↗
9. Third-body wear of cobalt–chromium–molybdenum implant alloys initiated by bone and poly(methyl methacrylate) particles10.1002/(SICI)1097-4636(20000605)50:3<322::AID-JBM5>3.0.CO;2-U ↗
10. Surface roughness of retrieved CoCrMo alloy femoral components from PCA artificial total knee joints10.1002/(SICI)1097-4636(2000)53:1<111::AID-JBM15>3.0.CO;2-Y ↗
11. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements10.1002/jor.1100170203 ↗
12. Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes (UHMWPE)10.1016/S0142-9612(99)00039-3 ↗
13. The radiation improvement of polyethylene prostheses: a preliminary study
14. Study of retrieved acetabular sockets made from high-dose, cross-linked polyethylene10.1054/arth.2001.28371 ↗
15. J-integral fracture toughness and tearing modulus measurement of radiation cross-linked UHMWPE10.1016/S0736-0266(02)00073-6 ↗
16. The effects of degree of crosslinking on the fatigue crack initiation and propagation resistance of orthopedic-grade polyethylene10.1002/jbm.a.10606 ↗
17. Mechanisms of decrease in fatigue crack propagation resistance in irradiated and melted UHMWPE10.1016/j.biomaterials.2005.06.025 ↗
18. Comparative fatigue behavior and toughness of remelted and annealed highly crosslinked polyethylenes10.1002/jbm.b.30807 ↗
19. Prevention of fatigue cracks in ultrahigh molecular weight polyethylene joint components by the addition of vitamin E10.1002/(SICI)1097-4636(1999)48:4<474::AID-JBM11>3.0.CO;2-T ↗
20. Characterization of irradiated blends of alpha-tocopherol and UHMWPE10.1016/j.biomaterials.2005.04.026 ↗
21. Alpha-tocopherol-doped irradiated UHMWPE for high fatigue resistance and low wear10.1016/j.biomaterials.2003.12.048 ↗
22. Comparative study of the performance of metallic and ceramic femoral head components in total replacement hip joints10.1016/0043-1648(96)80015-9 ↗
23. Hip-simulator ranking of polyethylene wear–comparisons between ceramic heads of different sizes10.3109/17453679608994656 ↗
24. Ceramic femoral heads prevent runaway wear for highly crosslinked polyethylene acetabular cups by third-body bone cement particles10.1016/S0043-1648(03)00289-8 ↗
25. Wear and creep of highly crosslinked polyethylene against cobalt chrome and ceramic femoral heads
26. Data evaluation in small-angle scattering–calculation of radial electron-density distribution by means of indirect Fourier transformation
27. A small-angle X-ray-scattering study of high-density polyethylene and ultrahigh molecular-weight polyethylene10.1021/ma00127a001 ↗
28. A study of the nanostructure and tensile properties of ultra-high molecular weight polyethylene10.1016/j.biomaterials.2003.10.027 ↗
29. The effect of counterface roughness on the wear of UHMWPE for rectangular wear paths10.1016/j.wear.2005.01.050 ↗
30. Quantification of the effect of cross-path motion on the wear rate of ultra-high molecular weight polyethylene10.1016/S0043-1648(03)00357-0 ↗
31. Tensile and tribological properties of high-crystallinity radiation crosslinked UHMWPE10.1002/jbm.b.31265 ↗
32. The combined effects of crosslinking and high crystallinity on the microstructural and mechanical properties of ultra high molecular weight polyethylene10.1016/j.biomaterials.2005.09.033 ↗
33. Advances in the processing, sterilization, and crosslinking of ultra-high molecular weight polyethylene for total joint arthroplasty10.1016/S0142-9612(99)00053-8 ↗
34. Polyethylene wear rate and osteolysis: critical threshold vs. continuous dose–response relationship10.1016/j.orthres.2004.11.005 ↗