journal article·2020-03-26Open access

A Simple Analytical Solution for the Designing of the Birdcage RF Coil Used in NMR Imaging Applications

Young Cheol Kim, Hyun Deok Kim, Byoung-Ju Yun, Sheikh Faisal Ahmad

Applied Sciences· Volume 10· Issue 7

Abstract

A novel analytical solution for the designing of the birdcage RF coil has been demonstrated in this paper. A new concept of dominant resonance path has been introduced in this paper which is used to identify the specific closed current loop in the birdcage RF coil which is responsible for the dominant resonance frequency mode. This concept is used to determine the precise numerical values of the lumped capacitance deployed in the legs and/or end-rings of the birdcage RF coil for its proper operation at the desired resonance frequency. The analytical solution presented in this paper has been established by performing the two-port network based equivalent circuit modeling of the birdcage RF coil. The proposed analytical solution uses T-matrix theory and develops a relationship between the input impedance of the birdcage coil and the impedances of its leg and end-ring segments. The proposed analytical solution provides the information about the resonance frequency spectrum of the birdcage RF coil and solves the issue of its interfacing with external circuits without affecting its resonance characteristics. Based upon the proposed analysis and designing strategy presented in this paper, the low pass, high pass and band pass configurations of the birdcage RF coil were successfully implemented with FPCB (Flexible Printed Circuit board) technique for small volume NMR imaging applications at 1.5 T and 3.0 T MRI system. The results obtained for the implemented birdcage coils using the proposed analysis and designing technique are in closed agreement with already established methods.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Advanced MRI Techniques and Applications · NMR spectroscopy and applications · Advanced NMR Techniques and Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Institute of Advanced Convergence Technology, Kyungpook National University, 80 Daehak-ro, Buk-gu, Daegu 41566, Korea

School of Electronics Engineering, College of IT Engineering, Kyungpook National University, 80 Daehak-ro, Buk-gu, Daegu 41566, Korea

Funding

Ministry of Trade, Industry and Energy · BK21 Plus project funded by the Ministry of Education Korea

References (26)

Show source-provided references
1. Imaging by nuclear magnetic resonance10.1088/0022-3735/21/1/002 ↗
2. Imaging by nuclear magnetic resonance and its bio-medical implications10.1016/0141-5425(79)90036-0 ↗
3. How to get ahead in imaging10.1038/458925a ↗
4. NMR probeheads for in vivo applications10.1002/1099-0534(2000)12:6<361::AID-CMR1>3.0.CO;2-L ↗
5. Resolution improvements in in vivo 1H NMR spectra with increased magnetic field strength10.1006/jmre.1998.1542 ↗
6. The development of the birdcage resonator: A historical perspective10.1002/nbm.1431 ↗
7. FPCB-based birdcage-type receiving coil sensor for small animal 1H 1.5T magnetic resonance imaging system
8. An efficient birdcage resonator at 2.5 MHz using a novel multilayer self-capacitance construction technique10.1007/BF01771702 ↗
9. Ahmad, S.F., Kim, Y.C., Choi, I.C., and Kim, H.D. (2015, January 1–4). Birdcage type NMR receiver coil sensor with integrated detuning circuit for 3T MRI system. Proceedings of the IEEE SENSORS 2015, Busan, Korea.10.1109/ICSENS.2015.7370638 ↗
10. Transceiver 4-leg birdcage for high field MRI: Knee imaging
11. A Hybrid Bird Cage Resonator for Sodium Observation at 4.7 T10.1002/mrm.1910240116 ↗
12. A practical multinuclear transceiver volume coil for in-vivo MRI/MRS at 7T10.1016/j.mri.2011.08.007 ↗
13. A birdcage coil tuned by RF shielding for applications at 9.4T10.1006/jmre.1997.1334 ↗
14. An efficient, highly homogeneous radiofrequency coil for whole-body NMR imaging at 1.5 T
15. Resonant modes of birdcage coil10.1006/jmre.1996.7488 ↗
16. The Theory of the bird-cage resonator
17. The Theory of an arbitrarily perturbed bird-cage resonator and a simple method for restoring it to full symmetry
18. Generalized electrical analysis of low-pass and high-pass birdcage resonators10.1016/0730-725X(91)90428-O ↗
19. A fast and accurate simulator for the design of birdcage coils in MRI10.1007/BF02693842 ↗
20. Advanced theory of driven birdcage resonator with losses for biomedical magnetic resonance imaging and spectroscopy10.1016/j.mri.2010.08.001 ↗
21. Ahmad, S.F., Kim, Y.C., Choi, I.C., Choi, S.W., Kim, T.G., Ahn, C.M., and Kim, H.D. (2017, January 4–6). Fast and efficient birdcage coil design process for high field MRI system by combining equivalent circuit model and 3D electromagnetic simulation. Proceedings of the 2017 Asia Modelling Symposium (AMS), Kota Kinabalu, Malaysia.10.1109/AMS.2017.37 ↗
22. Analysis of the bird-cage receiver coil for MRI system employing a equivalent circuit model based on transmission matrix
23. Jianming, J. (1999). Analysis and design of RF coils. Electromagnetic Analysis and Design in Magnetic Resonance Imaging, CRC Press. [1st ed.].
24. Homogeneous resonators for magnetic resonance: A review10.1016/j.crci.2007.10.003 ↗
25. BirdcageBuilder: Design of specified-geometry birdcage coils with desired current pattern and resonant frequency10.1002/cmr.10030 ↗
26. (2019, December 20). ANSYS Electronics Desktop. Available online: http://www.ansys.com/products/electronics/ansys-electronics-desktop.