journal article·2020-12-29Open access

Deep-Learning Based Positron Range Correction of PET Images

Joaquín L. Herraiz, Adrián Bembibre, Alejandro López-Montes

Applied Sciences· Volume 11· Issue 1

Abstract

Positron emission tomography (PET) is a molecular imaging technique that provides a 3D image of functional processes in the body in vivo. Some of the radionuclides proposed for PET imaging emit high-energy positrons, which travel some distance before they annihilate (positron range), creating significant blurring in the reconstructed images. Their large positron range compromises the achievable spatial resolution of the system, which is more significant when using high-resolution scanners designed for the imaging of small animals. In this work, we trained a deep neural network named Deep-PRC to correct PET images for positron range effects. Deep-PRC was trained with modeled cases using a realistic Monte Carlo simulation tool that considers the positron energy distribution and the materials and tissues it propagates into. Quantification of the reconstructed PET images corrected with Deep-PRC showed that it was able to restore the images by up to 95% without any significant noise increase. The proposed method, which is accessible via Github, can provide an accurate positron range correction in a few seconds for a typical PET acquisition.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Medical Imaging Techniques and Applications · Radiation Detection and Scintillator Technologies · Advanced Radiotherapy Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Nuclear Physics Group and IPARCOS, Faculty of Physical Sciences, University Complutense of Madrid, CEI Moncloa, 28040 Madrid, Spain

Health Research Institute of the Hospital Clínico San Carlos (IdISSC), 28040 Madrid, Spain

Funding

National Institutes of Health · Ministerio de Ciencia e Innovación · Consejería de Educación e Investigación

References (51)

Show source-provided references
1. Positron Emission Tomography: Current Challenges and Opportunities for Technological Advances in Clinical and Preclinical Imaging Systems10.1146/annurev-bioeng-071114-040723 ↗
2. Bailey, D.L., Townsend, D.W., Valk, P.E., and Maisey, M.N. (2005). Positron Emission Tomography: Basic Sciences, Springer.10.1007/b136169 ↗
3. Mettler, F.A., and Guiberteau, M.J. (2019). Essentials of Nuclear Medicine and Molecular Imaging, Elsevier.
4. (2020, November 26). PENELOPE-2018: A Code System for Monte Carlo Simulation of Electron and Photon Transport. Available online: https://www.oecd-nea.org/jcms/pl_46441/penelope-2018-a-code-system-for-monte-carlo-simulation-of-electron-and-photon-transport?details=true.
5. (2020, November 26). Progress in Positron Annihilation |Book| Scientific.Net. Available online: https://www.scientific.net/book/progress-in-positron-annihilation/978-3-03813-451-0.
6. Fundamental Limits of Spatial Resolution in PET10.1016/j.nima.2010.11.092 ↗
7. Calculation of positron range and its effect on the fundamental limit of positron emission tomography system spatial resolution10.1088/0031-9155/44/3/019 ↗
8. Study of CT-based positron range correction in high resolution 3D PET imaging10.1016/j.nima.2010.12.041 ↗
9. Positron range in PET imaging: An alternative approach for assessing and correcting the blurring10.1088/0031-9155/57/12/3931 ↗
10. Positron range estimations with PeneloPET10.1088/0031-9155/58/15/5127 ↗
11. The Impact of Positron Range on PET Resolution, Evaluated with Phantoms and PHITS Monte Carlo Simulations for Conventional and Non-conventional Radionuclides10.1007/s11307-019-01337-2 ↗
12. Radionuclide generators: The prospect of availing PET radiotracers to meet current clinical needs and future research demands
13. Tissue-Dependent and Spatially-Variant Positron Range Correction in 3D PET10.1109/TMI.2015.2436711 ↗
14. Improving PET Quantification of Small Animal [68Ga]DOTA-Labeled PET/CT Studies by Using a CT-Based Positron Range Correction10.1007/s11307-018-1161-7 ↗
15. Positron range in tissue-equivalent materials: Experimental microPET studies10.1088/0031-9155/61/17/6307 ↗
16. An overview of recent developments in FLUKA PET tools10.1016/j.ejmp.2018.06.636 ↗
17. Caribé, P.R.R.V., Vandenberghe, S., Diogo, A., Pérez-Benito, D., Efthimiou, N., Thyssen, C., D’Asseler, Y., and Koole, M. (2020). Monte Carlo Simulations of the GE Signa PET/MR for Different Radioisotopes. Front. Physiol., 11.10.3389/fphys.2020.525575 ↗
18. PenNuc: Monte Carlo simulation of the decay of radionuclides10.1016/j.cpc.2019.08.002 ↗
19. Bai, B., Ruangma, A., Laforest, R., Tai, Y.-, and Leahy, R.M. (2003, January 19–25). Positron range modeling for statistical PET image reconstruction. Proceedings of the 2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515), Portland, OR, USA.
20. Identification of a Shine-Through Artifact in the Trachea with 124I PET/CT10.2967/jnumed.108.060442 ↗
21. Bai, B., Laforest, R., Smith, A.M., and Leahy, R.M. (2005, January 23–29). Evaluation of MAP image reconstruction with positron range modeling for 3D PET. Proceedings of the IEEE Nuclear Science Symposium Conference Record, Fajardo, PR, USA.
22. Modeling and incorporation of system response functions in 3-D whole body PET10.1109/TMI.2006.873222 ↗
23. Alessio, A., and MacDonald, L. (2008, January 19–25). Spatially Variant Positron Range Modeling Derived from CT for PET Image Reconstruction. Proceedings of the 2008 IEEE Nuclear Science Symposium Conference Record, Dresden, Germany.10.1109/NSSMIC.2008.4774106 ↗
24. A residual correction method for high-resolution PET reconstruction with application to on-the-fly Monte Carlo based model of positron range10.1118/1.3284980 ↗
25. PET iterative reconstruction incorporating an efficient positron range correction method10.1016/j.ejmp.2015.11.005 ↗
26. Compressive effect of the magnetic field on the positron range in commonly used positron emitters simulated using Geant410.1140/epjp/i2017-11779-x ↗
27. Mathematical Removal of Positron Range Blurring in High Resolution Tomography10.1109/TNS.1986.4337166 ↗
28. Rukiah, A.L., Meikle, S.R., Gillam, J.E., and Kench, P.L. (2018, January 10–17). An investigation of 68Ga positron range correction through de-blurring: A simulation study. Proceedings of the 2018 IEEE Nuclear Science Symposium and Medical Imaging Conference Proceedings (NSS/MIC), Sydney, Australia.10.1109/NSSMIC.2018.8824301 ↗
29. PET reconstruction with system matrix derived from point source measurements10.1109/TNS.2005.862979 ↗
30. FIRST: Fast Iterative Reconstruction Software for (PET) tomography10.1088/0031-9155/51/18/007 ↗
31. Fast and efficient fully 3D PET image reconstruction using sparse system matrix factorization with GPU acceleration10.1088/0031-9155/56/20/015 ↗
32. Rahmim, A., Qi, J., and Sossi, V. (2013). Resolution modeling in PET imaging: Theory, practice, benefits, and pitfalls. Med. Phys., 40.10.1118/1.4800806 ↗
33. Reader, A.J., Corda, G., Mehranian, A., da Costa-Luis, C., Ellis, S., and Schnabel, J.A. (2020). Deep Learning for PET Image Reconstruction. IEEE Trans. Radiat. Plasma Med. Sci.10.1109/TRPMS.2020.3014786 ↗
34. Machine Learning in PET: From Photon Detection to Quantitative Image Reconstruction10.1109/JPROC.2019.2936809 ↗
35. Ronneberger, O., Fischer, P., and Brox, T. (2015). U-Net: Convolutional Networks for Biomedical Image Segmentation. arXiv.10.1007/978-3-319-24574-4_28 ↗
36. Herraiz, J.L. (2020, December 27). Deep PRC (2020)—Github Repository. Available online: https://github.com/jlherraiz/deepPRC.
37. A PENELOPE-based system for the automated Monte Carlo simulation of clinacs and voxelized geometries—Application to far-from-axis fields10.1118/1.3643029 ↗
38. A preclinical micro-computed tomography database including 3D whole body organ segmentations10.1038/sdata.2018.294 ↗
39. Performance evaluation of an Inveon PET preclinical scanner10.1088/0031-9155/54/9/020 ↗
40. Badal, A., Domarco, J., Udias, J.M., and Herraiz, J.L. (2018). MCGPU-PET: A Real-Time Monte Carlo PET Simulator, International Symposium on Biomedical Imaging.
41. Accelerating Monte Carlo simulations of photon transport in a voxelized geometry using a massively parallel graphics processing unit10.1118/1.3231824 ↗
42. GPU-Based Fast Iterative Reconstruction of Fully 3-D PET Sinograms10.1109/TNS.2011.2158113 ↗
43. Abadi, M., Barham, P., Chen, J., Chen, Z., Davis, A., Dean, J., Devin, M., Ghemawat, S., Irving, G., and Isard, M. (2016, January 2–4). TensorFlow: A System for Large-Scale Machine Learning. Proceedings of the OSDI’16: Proceedings of the 12th USENIX conference on Operating Systems Design and Implementation, Savannah, GA, USA.
44. Chollet, F. (2020, November 25). Others Keras. Available online: https://keras.io.
45. Berker, Y., Maier, J., and Kachelrieß, M. (2018, January 10–17). Deep Scatter Estimation in PET: Fast Scatter Correction Using a Convolutional Neural Network. Proceedings of the 2018 IEEE Nuclear Science Symposium and Medical Imaging Conference Proceedings (NSS/MIC), Sydney, Australia.10.1109/NSSMIC.2018.8824594 ↗
46. Leung, K.H., Marashdeh, W., Wray, R., Ashrafinia, S., Rahmim, A., and Jha, A.K. (2020). A Physics-Guided Modular Deep-Learning Based Automated Framework for Tumor Segmentation in PET Images. arXiv.10.1088/1361-6560/ab8535 ↗
47. Ramachandran, P., Zoph, B., and Le, Q.V. (2017). Swish: A Self-Gated Activation Function. arXiv.
48. Liu, L., Jiang, H., He, P., Chen, W., Liu, X., Gao, J., and Han, J. (2020). On the Variance of the Adaptive Learning Rate and Beyond. arXiv.
49. Zhang, M.R., Lucas, J., Hinton, G., and Ba, J. (2019). Lookahead Optimizer: K steps forward, 1 step back. arXiv.
50. Kuijper, A., Bredies, K., Pock, T., and Bischof, H. (2013). Outlier Removal Power of the L1-Norm Super-Resolution. Lecture Notes in Computer Science, Springer. Scale Space and Variational Methods in Computer Vision; SSVM 2013.10.1007/978-3-642-38267-3 ↗
51. Feasibility study of the positronium imaging with the J-PET tomograph10.1088/1361-6560/aafe20 ↗